Model Number 1707801
Device Problems Low Test Results (2458); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/02/2023
Event Type  malfunction  
Event Description
A customer contacted the ortho clinical diagnostics (ortho) technical solution center (tsc) to report a non-reproducible, lower than expected glucose (glu) result was obtained from a single patient sample processed using vitros chemistry products glu slides lot 0019-3227-0206 on a vitros 4600 chemistry system.Patient sample result of < 20 mg/dl vs an expected result of 193 mg/dl biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected.The lower than expected vitros glu result was not reported from the laboratory.Ortho has not been made aware of any allegation of patient harm as a result of this event.This report corresponds to ortho clinical diagnostics inc (ortho) complaint number (b)(4)and reportability assessment (b)(4).
 
Manufacturer Narrative
The investigation has determined that a non-reproducible, lower than expected glucose (glu) result was obtained from a single patient sample processed using vitros chemistry products glu slides lot 0019-3227-0206 on a vitros 4600 chemistry system.A definitive assignable cause of the event could not be determined.Based on historical quality control results, a vitros glu lot 0019-3227-0206 performance issue cannot be ruled out as a contributor to the event.As no precision testing was performed on the vitros 4600 chemistry system, an instrument related issue cannot be entirely ruled out as a contributor of the event.However, there was no indication that the analyzer malfunctioned, and historical qc results were precise indicating that an issue related to the performance of the vitros 4600 system was not a likely contributing factor of the event.Pre-analytical sample processing could not be ruled out as a contributing factor as it is unknown if the customer was following the sample collection device manufacturer¿s recommended centrifugation protocol.Improper pre-analytical sample handling could have contributed to the event.It is possible that cellular debris, due to poor sample preparation, was present in the affected sample, although this could not be confirmed.Continual tracking and trending of complaint data has not identified any signals to suggest there is a systemic quality issue with vitros glu reagent lot 0019-3227-0206.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITROS CHEMISTRY PRODUCTS GLU SLIDES
Type of Device
IN-VITRO DIAGNOSTICS
Manufacturer (Section D)
ORTHO-CLINICAL DIAGNOSTICS, INC.
100 indigo creek drive
rochester NY 14626
Manufacturer (Section G)
ORTHO-CLINICAL DIAGNOSTICS, INC.
513 technology boulevard
rochester NY 14626
Manufacturer Contact
laurie o'riordan
microtyping systems
1295 southwest 29th avenue
pompano beach, FL 33069
9549709500
MDR Report Key17230969
MDR Text Key318142052
Report Number0001319809-2023-00053
Device Sequence Number1
Product Code CGA
UDI-Device Identifier10758750009572
UDI-Public10758750009572
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/01/2023
Device Model Number1707801
Device Catalogue Number1707801
Device Lot Number0019-3227-0206
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/06/2023
Initial Date FDA Received06/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-