G3: the aware date for this event is 06/05/2023 based off of new information provided.Prismaflex sepxiris set 150 (jp) is similar to prismaflex st150 set, prismaflex st150 set ckt, and prismaflex st150 set c.Prismaflex st150 set, prismaflex st150 set ckt, and prismaflex st150 set c have been temporarily approved for use in the us under emergency use authorization eua(b)(4) to deliver crrt to treat patients in an acute care environment during the covid-19 pandemic.H10: the device was received for evaluation as well as a video sample.The video was reviewed, and it was noted that there was a leak from the warmer connector.Visual inspection of the actual sample did not identify any abnormalities that could have contributed to the reported condition.A functional air leak test was also performed, and no issues were noted.The reported condition was verified during the video sample review.The cause of the reported condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|