• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION SEPXIRIS SET; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION SEPXIRIS SET; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955617
Device Problems Fluid/Blood Leak (1250); Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2023
Event Type  malfunction  
Manufacturer Narrative
G3: the aware date for this event is 06/05/2023 based off of new information provided.Prismaflex sepxiris set 150 (jp) is similar to prismaflex st150 set, prismaflex st150 set ckt, and prismaflex st150 set c.Prismaflex st150 set, prismaflex st150 set ckt, and prismaflex st150 set c have been temporarily approved for use in the us under emergency use authorization eua(b)(4) to deliver crrt to treat patients in an acute care environment during the covid-19 pandemic.H10: the device was received for evaluation as well as a video sample.The video was reviewed, and it was noted that there was a leak from the warmer connector.Visual inspection of the actual sample did not identify any abnormalities that could have contributed to the reported condition.A functional air leak test was also performed, and no issues were noted.The reported condition was verified during the video sample review.The cause of the reported condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that loose connection occurred between the luer connector - warmer connection on the return circuit of a prismaflex sepxiris set 150 which resulted in an external fluid leak.This issue occurred while priming the set prior to patient use.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEPXIRIS SET
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
av lionel terray, b.p. 126
7
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17233373
MDR Text Key318215065
Report Number8010182-2023-00239
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955617
Device Lot Number22G0032
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/05/2023
Initial Date FDA Received06/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-