MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM X; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
|
Back to Search Results |
|
Model Number 97800 |
Device Problems
Failure to Deliver Energy (1211); Energy Output Problem (1431); Ambient Noise Problem (2877); Application Program Problem (2880)
|
Patient Problem
Discomfort (2330)
|
Event Date 05/31/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a patient regarding an implantable neurostimulator for the treatment of bladder issues.It was reported that about 4 -5 days ago noticed a change in stimulation sensation, said that could feel the vibration whereas before didn't feel anything.When asked, patient said that they did have an ultrasound at the hospital and said that the ultrasound was done on their stomach.Patient misplaced their communicator and can't make an adjustment.Replacement request was sent to repair.The patient was redirected to their healthcare provider to further address the issue.Replacement request was sent to repair.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received.The patient called back and reports that they found the communicator that they thought was lost.Caller was trying to use it and seeing "communicator not found" helped caller use the "switch communicator" function and pair the communicator.The caller reports that they are seeing the max settings reached screen at 0.4.The caller reports no falls or trauma.The caller tried to change programs and encountered the same message.Reviewed that they can try the rest of the programs, but they should make an appointment with the hcp.The caller noted that they have been seeing this maximum settings screen starting 2 days ago.
|
|
Search Alerts/Recalls
|
|
|