• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MULTI MEASUREMENT SERVER X2

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MULTI MEASUREMENT SERVER X2 Back to Search Results
Model Number M3002A
Device Problem Inaudible or Unclear Audible Prompt/Feedback (2283)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/27/2022
Event Type  malfunction  
Manufacturer Narrative
A philips remote service engineer (rse) remotely interviewed the customer who was onsite.The field service engineer (fse) confirmed the sound was distorted in standalone mode.However, the customer reported that they noticed the issue in the inoperative message as they don¿t use it in stand alone mode.Based on the information available, the reported problem was confirmed.After speaker replacement the device was returned to functional use with no further issues identified.The device remains at the customer site.No further investigation or action is warranted at this time.
 
Event Description
Philips received a complaint on the intellivue multi measurement server x2 sn (b)(6) indicating a speaker operation error message was displayed.The device was in use monitoring a patient at the time of the event.No adverse event involving a patient or user was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MULTI MEASUREMENT SERVER X2
Type of Device
INTELLIVUE MULTI MEASUREMENT SERVER X2
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key17238862
MDR Text Key318278360
Report Number9610816-2023-00321
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838000261
UDI-Public00884838000261
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K150310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3002A
Device Catalogue NumberM3002A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/27/2022
Initial Date FDA Received06/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/17/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-