|
Model Number CD2411-36Q |
Device Problems
Failure to Deliver Shock/Stimulation (1133); Failure to Convert Rhythm (1540); Low impedance (2285)
|
Patient Problems
Arrhythmia (1721); Syncope/Fainting (4411)
|
Event Date 01/07/2023 |
Event Type
Injury
|
Event Description
|
Related manufacturer reference numbers: 2017865-2023-24289.It was reported that the patient presented in clinic for a follow-up.Upon interrogation, it was found that the implantable cardioverter defibrillator (icd) and right ventricular (rv) lead had failed to deliver a high voltage therapy with low defibrillation impedance exhibited, and noise was present on the rv lead.It was also noted that the patient had fallen and had hip surgery around that time.The icd was explanted and replaced, and programming changes were made for the rv lead.The patient was in stable condition pre, during, and post procedure.
|
|
Manufacturer Narrative
|
The reported events of low impedance, failure to deliver shock, and failure to convert rhythm were confirmed in the device image by tech services but was not due to a device anomaly.The device behaved as programmed.The device was above elective replacement indicator (eri) when received.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.The anomaly observed in the field could not be reproduced.
|
|
Search Alerts/Recalls
|
|
|