MAUDE Adverse Event Report: MEDTRONIC IRELAND VENA SEAL CLOSURE SYSTEM; AGENT, OCCLUDING, VASCULAR, PERMANENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND VENA SEAL CLOSURE SYSTEM; AGENT, OCCLUDING, VASCULAR, PERMANENT Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pulmonary Embolism (1498); Granuloma (1876); Phlebitis (2004); Local Reaction (2035)
Event Type  Injury  
Manufacturer Narrative
Abstract title 'long-term practical experience with cyanoacrylate'.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received an abstract titled 'long-term practical experience with cyanoacrylate'.Between january 2014 and january 2022, a total of 687 patients were treated with the cyanoacrylate method (venaseal procedure) at the center for vascular medicine düsseldorf in germany.During the same period, 1762 veins were treated by thermal methods such as radiofrequency or laser and respectively in 1380 by conventional open crossectomy and stripping.Patients were followed up retrospectively in a follow-up of up to 36 months in terms of success rate and complications.It was often that a single vein was treated-87%, namely more ssv than gsv.Patients have always had chronic venous disease st.Ii and iii (ceap c2-c6).63 x allergic reactions (9.1%), 35 x phlebitides (5%), 3 foreign body granulomas and a pulmonary artery embolism have been documented.The primary closure was 100%.Partial recanalization (2.5%) was documented in 17 patients and complete recanalization (0.3%) in a further 2 patients.After 12 months, 26 partial sewerage systems (3.8%) and 3 complete re-sewers (0.44%) were detected in total.The closure rate after 36 months was 95.7%.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENA SEAL CLOSURE SYSTEM
Type of Device
AGENT, OCCLUDING, VASCULAR, PERMANENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17245354
MDR Text Key318343029
Report Number9612164-2023-02935
Device Sequence Number1
Product Code PJQ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P140018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Literature,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/19/2022
Initial Date FDA Received07/03/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-