INTEGRA LIFESCIENCES (IRELAND) LTD. DUO HEADLIGHT 2 BAY SYSTEM - US; PFM05 Back to Search Results
Model Number 90620US
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The duo headlight 2 bay system - us (90620us) was returned for evaluation: failure analysis: the returned duo headlight was received in used condition.Evaluation could not identify any issues with the headlight overheating or flickering; however, evaluation did identify that the fan was loud while the unit was tested.The fan has been replaced and the unit was tested and passed all the testing.Root cause analysis: the reported complaint could not be confirmed.The issue of the unit overheating and flickering could be due to a malfunctioning fan.No manufacturing, workmanship, or material deficiency has been identified.
 
Event Description
It was reported that the cooling fan of the duo headlight 2 bay system - us (90620us) operated intermittently, causing the light to overheat & flicker.It is unknown under what circumstance this event occurred; however, no patient injury, death or surgical delay was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO HEADLIGHT 2 BAY SYSTEM - US
Type of Device
PFM05
Manufacturer (Section D)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI 
Manufacturer (Section G)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key17247220
MDR Text Key318551255
Report Number3006697299-2023-00070
Device Sequence Number1
Product Code FSR
UDI-Device Identifier10381780490746
UDI-Public10381780490746
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 07/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number90620US
Device Catalogue Number90620US
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2023
Initial Date Manufacturer Received 06/12/2023
Initial Date FDA Received07/03/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-