Model Number N/A
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/07/2023
Event Type  malfunction  
Event Description
It was reported that the femoral stem was not correlating with broach size.The surgeon believed it was oversized from plasma spray.There was no known impact or consequences to the patient.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
(b)(4).D10: cat# 010000663 lot# 7435143 g7 pps ltd acet shell 52e.Cat# 650-0661 lot# 3095707 delta ceramic fem hd 36/0mm.Cat# 30103605 lot# 65055789 g7 vit e neutral lnr 36mm e.The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected}.Mechanical (g04) ¿ stem.A tprlc was returned and evaluated against the complaint.Visual inspection confirmed the etching on the part to match the provided lot and size description.Dings were observed around the stem inserter hole.Foreign debris remains affixed to the coating in multiple locations over the surface.Calipers were used to measure the product identifiers for the stem.The thickness was measured in the upper limit of the specification.The overall length was also measured and found to be within the stated print tolerance.A review of the device history records identified no related deviations or anomalies during manufacturing.Medical records were not provided.The investigation could not verify or identify any evidence of product contribution to the reported problem.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further information at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 T1 PPS SO 11X142MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17248753
MDR Text Key318380024
Report Number0001825034-2023-01470
Device Sequence Number1
Product Code LZO
UDI-Device Identifier00880304489745
UDI-Public(01)00880304489745(17)320803(10)7303264
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200196
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-103110
Device Lot Number7303264
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/07/2023
Initial Date FDA Received07/03/2023
Supplement Dates Manufacturer Received10/10/2023
Supplement Dates FDA Received10/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-