MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH ORIGINAL M.E. MÜLLER®, STEM, PT-10, STRAIGHT, LATERAL, CEMENTED,15.0,TAPER 12/14; HIP PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH ORIGINAL M.E. MÜLLER®, STEM, PT-10, STRAIGHT, LATERAL, CEMENTED,15.0,TAPER 12/14; HIP PROSTHESIS Back to Search Results
Model Number N/A
Device Problem Patient Device Interaction Problem (4001)
Patient Problems Failure of Implant (1924); Pain (1994); Osteolysis (2377); Swelling/ Edema (4577)
Event Date 05/26/2023
Event Type  Injury  
Manufacturer Narrative
(b)(6).G2: foreign - germany d10: item name: metasulâ® ldhâ®, head, 50, code p, taper 18/20; item reference: 01.00181.500; item lot: 2307129 item name: metasul ldh head adapter ¿m¿/0, taper 12/14-18/20; item reference: 01.00185.146; item lot: 2345877 item name: durom acetabular component 56/50p; item reference: 01.00214.056; item lot: 2325907 multiple mdr reports were filed for this event, please see associated reports: 0009613350-2023-00352 0009613350-2023-00353 0009613350-2023-00355 investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that patient underwent revision of total hip implant due to pain, swelling and redness in the scar area.During the revision it was noted that the patient has a extraperitoneal infection with metal abrasion in the secretion.Other findings are: osteolysis, metal abrasion ring around the stem and detachment of titanium coating from acetabular cup.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.No product was returned, however pictures of the stem and the cup were provided and assessed.Only the proximal part of the stem is visible.The stem is abraded in the are between the stem taper and the neck of the stem.Black tissue is visible in the surgical area of the hip.Only the outer part of the cup is visible.The coating of the cup is delaminated and most likely came off during removal of the cup from the acetabulum during revision surgery.Due to the low quality of the pictures a more in depth assessment is not possible.A review of the device manufacturing records confirmed no abnormalities or deviations.Devices are used for treatment medical records were not provided.With the available information, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORIGINAL M.E. MÜLLER®, STEM, PT-10, STRAIGHT, LATERAL, CEMENTED,15.0,TAPER 12/14
Type of Device
HIP PROSTHESIS
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key17253682
MDR Text Key318350590
Report Number0009613350-2023-00354
Device Sequence Number1
Product Code KWA
UDI-Device Identifier00889024592742
UDI-Public(01)00889024592742(17)101031(10)2296543
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/31/2010
Device Model NumberN/A
Device Catalogue Number12.00.39-150
Device Lot Number2296543
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/05/2023
Initial Date FDA Received07/04/2023
Supplement Dates Manufacturer Received07/18/2023
Supplement Dates FDA Received07/31/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/28/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10.
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexPrefer Not To Disclose
-
-