Model Number HAMILTON-C3
Device Problems Suction Problem (2170); Pressure Problem (3012)
Patient Problem Unspecified Respiratory Problem (4464)
Event Date 05/24/2021
Event Type  Injury  
Event Description
Ventilator alarmed with tf341009, tf346054, te 231036 and switched to safety mode 385002.
 
Manufacturer Narrative
The complaint has been reopened and reviewed according to fda form 483 inspectional observation ems #2, eobs2 from the fda inspection conducted between july 17 to july 21, 2022 at the ems and bonaduz sites.  a detailed investigation was performed by an expert from the technical service: this issue is deemed a reportable event since the malfunction "switch on safety mode due to tf 341009" can lead to a delay in the therapy since the ventilator cannot be used.Eventually a new ventilator needs to be prepared.The root cause of the switch on safety mode was determined to be an unintended use of the suctioning tool in conjunction with a closed suctioning system having as effect an incorrect interpretation of the situation.The correction in this case was to strictly follow the instruction for use.With this investigation it has been confirmed that the device failed to meet its specifications at the time of theevent, while the medical device was used for treatment (in ventilation phase).There was no patient or user harm.The allegation in this complaint was confirmed to be a complaint.No further investigation or correction will be performed except those mentioned above.In future hamilton medical ag will report an event like this issue as it will be deemed a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAMILTON MEDICAL AG
Type of Device
HAMILTON-C3
Manufacturer (Section D)
HAMILTON MEDICAL AG
via crusch 8
bonaduz, graubuenden 7402
SZ  7402
Manufacturer Contact
nikolai sonnenberg
via crusch 8
bonaduz, 7402
SZ   7402
MDR Report Key17253684
MDR Text Key318350600
Report Number3001421318-2023-22775
Device Sequence Number1
Product Code CBK
UDI-Device Identifier07630002808590
UDI-Public07630002808590
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K201306
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberHAMILTON-C3
Device Catalogue Number161005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/29/2021
Initial Date FDA Received07/04/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-