• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ISOFLEX OPTIM LEAD; Permanent defibrillator electrodes

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ISOFLEX OPTIM LEAD; Permanent defibrillator electrodes Back to Search Results
Model Number 1948/58
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/06/2023
Event Type  malfunction  
Manufacturer Narrative
The reported event of threshold was not confirmed.As received, a complete lead was returned in one piece for analysis.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual and x-ray inspections of the lead did not find any anomalies except for procedural damages.
 
Event Description
It was reported that the patient presented for an implant procedure.It was noted that right ventricular lead failed to sense the threshold even after placing in several positions.The lead was not used and replaced during procedure.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISOFLEX OPTIM LEAD
Type of Device
Permanent defibrillator electrodes
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17253862
MDR Text Key318350882
Report Number2017865-2023-24981
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734502351
UDI-Public05414734502351
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P960030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 07/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1948/58
Device Catalogue Number1948-58
Device Lot NumberP000159669
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/21/2023
Initial Date FDA Received07/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/13/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-