Catalog Number 8065752914
Device Problem Excess Flow or Over-Infusion (1311)
Patient Problem Capsular Bag Tear (2639)
Event Date 06/12/2023
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that during a surgery while using a phacoemulsification handpiece and the fluidics management system (fms) pack there was a surge in perfusion and patient experienced capsule rupture.The surgery was completed without product replacement, and vitrectomy treatment for capsular rupture was performed by using another console.
 
Manufacturer Narrative
The product under investigation is not a serviceable device.Therefore, a service record review was not performed.However, a phaco handpiece (hp) was returned for testing on this investigation.A non-conformance-based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product /serial under investigation.The hp was received for testing on this investigation.A visual assessment of the returned sample revealed no nonconformities found.The hp was connected to a calibrated centurion.The handpiece was connected to dynamic tuning fixture (dtf) for stroke length testing on the longitudinal and torsional movements which found the handpiece to meet specifications.The handpiece was found to meet specifications.Therefore, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, ACCESSORY, ACTIVE SENTRY HANDPIECE
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17255117
MDR Text Key318373794
Report Number2028159-2023-00943
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K161794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065752914
Device Lot Number156YUN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/11/2023
Initial Date FDA Received07/05/2023
Supplement Dates Manufacturer Received09/12/2023
Supplement Dates FDA Received10/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CENTURION SURGICAL PROCEDURE PAK; CENTURION VISION SYSTEM
Patient Outcome(s) Other; Required Intervention;
-
-