MAUDE Adverse Event Report: KAVO DENTAL GMBH MASTERTORQUE LUX M8900 L; DENTAL HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KAVO DENTAL GMBH MASTERTORQUE LUX M8900 L; DENTAL HANDPIECE Back to Search Results
Model Number M8900 L
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/06/2023
Event Type  malfunction  
Manufacturer Narrative
The instrument was checked visually and functionally against the valid specification.The retention force of the chuck system was tested, and it was within specification.The complained issue was not reproducible.Hence it has to be assumed that either the used bur did not meet the specification or it was not correctly inserted.To avoid such issues the user instruction contains already several notes, warnings and requests how to prepare and how to check the handpiece for each treatment and how to use it.2.3 technical condition: a damaged device or components could injure patients, users and third parties.Only operate devices or components if they are undamaged on the outside.Check that the device is working properly and is in satisfactory condition before each use.Have parts with sites of breakage or surface changes checked by the service.If the following defects occur, stop working and have the service personnel carry out repair work: malfunctions, damage, irregular running noise, excessive vibration, overheating, dental bur is not seated firmly in the handpiece.Observe the following instructions in order to guarantee optimum functioning and prevent material damage: service the medical device with care products and systems regularly as described in the instructions for use.The device should be reprocessed and stored in a dry location, according to instructions, if it is not be used for a longer period.2.6 service and repair: repairs, servicing and safety checks may only be performed by trained service personnel.The following persons are authorized to do this: service technicians of kavo branches after the appropriate product training, service technicians of kavo authorized dealers after the appropriate product training.Observe all the following items during servicing work: have the service and testing tasks carried out according to the medical device operator ordinance.After servicing, interventions on and repairs of the device and before re-use, have the service personnel perform safety checks on the device.Following expiry of the warranty, have the tool holding system checked once a year.Kavo recommends specifying in-house service intervals where the medical device is brought to a professional shop for cleaning, servicing and a function check.Define the service interval depending on the frequency of use.As a result of the use of non-kavo original spare parts during the repair, parts such as covers may become undone and injure the patient, user or other people.This may result in aspiration, swallowing of parts and possibly even a risk of suffocation.Only use spare parts that comply with the specification for repair; original kavo spare parts comply with the specification.Note: if a repair is done with non-kavo original spare parts, this may constitute a product modification that leads to the loss of ce conformity.In the event of damage, the responsibility is with the service company or the operator.The introduction into the market of a modified product, where there is reasonable suspicion that the safety and health of patients or users may be jeopardized, is prohibited by the german medical device law §4, section 1 no.1 and requires a separate conformity check.Caution: defective clamping system.Risk of injury, dental bur may fall out during treatment.Pull on the dental bur to check if the clamping system works properly and if the dental bur is firmly clamped.
 
Event Description
The dental office informed that during a dental treatment on tooth #31 the bur separated from handpiece and was swallowed by the patient.Bur passed the body on natural way without any problems.There was no medical care necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MASTERTORQUE LUX M8900 L
Type of Device
DENTAL HANDPIECE
Manufacturer (Section D)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, 88400
GM  88400
Manufacturer (Section G)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, 88400
GM   88400
Manufacturer Contact
klaus reisenauer
bismarckring 39
biberach / riss, 88400
GM   88400
MDR Report Key17256288
MDR Text Key319130369
Report Number3003637274-2023-00014
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130560
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Dentist
Type of Report Initial
Report Date 07/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM8900 L
Device Catalogue Number1.008.7800
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/15/2023
Initial Date FDA Received07/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/11/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-