BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number 5C4482 |
Device Problems
Failure to Disconnect (2541); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/05/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that there was a separation between the female connector (dark blue) and the main body of the minicap transfer set.The patient was instructed by their pd nurse to clamp the patient line twice from the amia cartridge and cut in-between and go into the clinic to have the transfer set changed.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Correction to h6 (medical device problem codes): remove a1204.H10: the actual device was not available; however, two (2) photographs of the sample were provided for evaluation.A visual inspection with the naked eye identified a separation between the female connector and main body on the transfer set.The reported condition was verified.The cause of the condition was determined to be manufacturing related issue due to an inadequate solvent bond between the female connector, insert chip, and main body of the twist clamp.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Search Alerts/Recalls
|
|
|