Catalog Number 8065753057 |
Device Problem
No Flow (2991)
|
Patient Problems
Corneal Scar (1793); Eye Burn (2523); Fluid Discharge (2686); Intraocular Pressure Decreased (4468); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 06/08/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported that during cataract surgery, a 64 years female patient experienced eye burn during phacoemulsification due to occlusion in handpiece and lack of saline flow it resulted in corneal opacification.There was minimal collapsing of the anterior chamber and noticed wound leak it was sutured.Procedure was completed and patient current condition was unknown.
|
|
Manufacturer Narrative
|
Service history was reviewed for the system.No service record relevant to the complaint reported event was found.However, the system was last serviced prior to the reported event per service record (sr) opened, for installation of the system.The system found to meet all cosmetic and performance a non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|