Model Number UHI-4
Device Problem Restricted Flow rate (1248)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/04/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation.During the evaluation, the customer's original reported issue of insufficient pressure was not confirmed.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
A user facility reported to olympus insufficient pressure on the high flow insufflation unit for a radical treatment of rectal cancer.There was no delay, and the patient was not under sedation.The procedure was completed with the same one; the malfunction was found during reprocessing.
 
Manufacturer Narrative
Updated fields: h6 and h10 this report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation the specific cause of phenomenon could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17262068
MDR Text Key318602620
Report Number3002808148-2023-06719
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/05/2023
Initial Date FDA Received07/05/2023
Supplement Dates Manufacturer Received07/19/2023
Supplement Dates FDA Received08/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/08/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-