A male patient with symptomatic benign prostatic hyperplasia (bph) received an aquablation procedure.Procept biorobotics corporation (procept) became aware that during the alignment phase, the aquabeam handpiece (hp) waterjet exhibited irregular jetting and the aquabeam cpu screen would intermittently freeze.While the jet worked as intended during the priming, the surgeon felt that the hp and trus were inadequately rotated based on the patient's positioning on the bed.As a result, the treating surgeon chose to abort the procedure without opting to try a new hp unit.There were no adverse health consequences to the patient due to this event.
|
The aquabeam handpiece was returned for investigation.Visual inspection of the returned device observed to have no physical damages or anomalies.Additionally, no debris was observed along the telescoping tube.Functional testing confirmed the reported event as the scope could not be inserted into the handpiece as the telescoping tube tube was stuck in place.Slight force had to applied to adjust the telescoping tube and and allow it to move properly.Root cause is undeterminable as the scope carriage could be retracted once the telescoping tube was allowed to move freely, and the failure could not be consistently produced.A review of the device history record (dhr) ab2000-b / serial number (b)(6) and aquabeam handpiece / lot number 23c01170 was conducted, which confirmed that there were no non-conformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system met all design and manufacturing specifications when released for distribution.The aquabeam robotic system user manual, um0101 rev.F, states the following: 11.2.5 sterile: aquabeam handpiece and aquabeam scope setup using the knobs on the carriage, advance the aquabeam scope forwards (distal) and backwards (proximal) to ensure that the aquabeam scope is fully engaged with the aquabeam handpiece.The scope tube tip should move forwards and backwards in concert with the movement of the aquabeam scope.The current instruction for use ifu0101-00 rev.E, aquabeam robotic system ifu, us, english was reviewed.8.30 sterile: treatment a.To begin the aquablation treatment, step on the foot pedal and gently support the beige braided tubing extending from the back of the aquabeam handpiece.Caution (for system models other than ab2000c): failure to support the beige braided tubing may result in a system fault or insufficient cutting efficacy.A.During the aquablation treatment, use [-] indicator on the motorpack to decrease power/resection depth (if needed) and use [+] to increase back to planned power submission of this report does not constitue an admission that the manufacturer's product caused or contributed to the event.
|