A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 6 years since the device was manufactured.The device was returned to an olympus repair facility and an evaluation was performed.Upon inspection and testing of the unit, the reported problem of noise could not be reproduced or confirmed.Additionally, the bending section cover had a scratch and a chip, and the number of broken wires in the bending tube blade exceeded the standard value, caused by mishandling.Switch 3 had a scratch, and the video connector had a crack.Based on the results of the investigation, a root cause of the reported event could not be determined.This issue is addressed in the instructions for use (ifu): ¿chapter 3 preparation and inspection, 3.8 inspection of the endoscopic system¿ describes the following warning. confirm that the wli and nbi endoscopic images are normal.1.Before inspection, wipe the objective lens using clean lint-free cloths moistened with saline solution or sterilized water.2.Observe the palm of your hand in the wli and nbi endoscopic images.3.Confirm that light is output from the endoscope¿s distal end.4.Adjust the brightness level as appropriate.5.Confirm that the wli and nbi endoscopic images are free from noise, blur, fog, or other irregularities.6.Turn the angulation control levers slowly in each direction until it stops.7.Confirm that the wli and nbi endoscopic images do not momentarily disappear or display any other irregularities.Olympus will continue to monitor the field performance of this device.
|