C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP MRI 6F CHRONW/OS; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 8806061 |
Device Problem
Device Contaminated During Manufacture or Shipping (2969)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/04/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.H10: d4 (expiry date: 02/2023).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Event Description
|
It was reported that prior to a port placement procedure in the right internal jugular vein, upon opening of the package, a little bug was found inside and the product was contaminated.There was no patient contact.
|
|
Manufacturer Narrative
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation.Two electronic photos were provided for review.The photo shows a partial view of one unsealed sample tray containing one dilator and a tunneler.A small pale foreign material was noted on the package in both the photos.Manufacturing site evaluation revealed a foreign matter inside the packaging, located near the tunnel and vessel dilator.Therefore, the investigation is confirmed for the reported foreign material contamination.The definitive root cause could not be determined based upon the available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: b5, d4 (expiration date: 02/2023), g3, h6 (method) h11: h6 (result, conclusion) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Event Description
|
It was reported that prior to a port placement procedure in the right internal jugular vein, upon opening of the package, a little bug was found inside and the product was contaminated.The procedure was completed using another device.There was no patient contact.
|
|