• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ELEKTA SOLUTIONS AB MOSAIQ; ACCELERATOR, LINEAR, MEDICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ELEKTA SOLUTIONS AB MOSAIQ; ACCELERATOR, LINEAR, MEDICAL Back to Search Results
Model Number MOSAIQ
Device Problem Computer Software Problem (1112)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/09/2023
Event Type  malfunction  
Manufacturer Narrative
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
 
Event Description
The customer reported that a pop up for table shifts is no longer displayed on the sequencer.
 
Manufacturer Narrative
H2 updated h6 updated.H10 updated.The investigation was completed by conducting a thorough evaluation of the product and the reported information.The investigation found from the log files that the cma window did display on the mosaiq/sequencer machine accordingly after the image registration was completed on the xvi machine.From the review, the cma window was displayed three times for the three times that mosaiq received the info from xvi.Based on available information there was no mistreatment.Mosaiq did not have any malfunction and was working as designed and intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIQ
Type of Device
ACCELERATOR, LINEAR, MEDICAL
Manufacturer (Section D)
ELEKTA SOLUTIONS AB
kungstensgatan 18
box 7593
stockholm, SE-10 3 93
SW  SE-103 93
Manufacturer (Section G)
ELEKTA SOLUTIONS AB
400 perimeter center terrace
suite 50
atlanta GA 30346
Manufacturer Contact
cornerstone
london road
crawley, west sussex RH10 -9BL
MDR Report Key17265062
MDR Text Key318872223
Report Number3015232217-2023-00040
Device Sequence Number1
Product Code IYE
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K183034
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberMOSAIQ
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/09/2023
Initial Date FDA Received07/06/2023
Supplement Dates Manufacturer Received06/09/2023
Supplement Dates FDA Received09/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-