• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC FABRICATION S.A.S. STREAMLINE TEMPORARY PACING LEADS; ELECTRODE, PACEMAKER, TEMPORARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC FABRICATION S.A.S. STREAMLINE TEMPORARY PACING LEADS; ELECTRODE, PACEMAKER, TEMPORARY Back to Search Results
Model Number 6495F
Device Problem Break (1069)
Patient Problem Insufficient Information (4580)
Event Date 05/04/2023
Event Type  malfunction  
Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that one day post implant of this temporary pacing lead, it was reported that when the dressing and a green needle was removed, it was observed that the lead had snapped into two portions about 2 inches from the skin.It was thought that "there was just enough of the exposed wire from that plastic casing, to attempt inserting it into the pacing connection if needed." both portions of the product were removed immediately and it was stated that no intervention was required as a result of the pacing lead snapping.It was mentioned that the lead attached to the pacemaker was connected to the patient by a green needle and one pacing wire.It was stated that there was the "green needle present and in the external pacing wire to complete the circuit for ventricular pacing." it was noted that staff are familiar in the use of this model of pacing lead.It was stated that the lead had been inspected prior to use with no issues noted.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STREAMLINE TEMPORARY PACING LEADS
Type of Device
ELECTRODE, PACEMAKER, TEMPORARY
Manufacturer (Section D)
MEDTRONIC FABRICATION S.A.S.
103 route d anor
fourmies nord 59610
FR  59610
Manufacturer (Section G)
MEDTRONIC FABRICATION S.A.S.
103 route d anor
fourmies nord 59610
FR   59610
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key17268849
MDR Text Key318863576
Report Number9611350-2023-00007
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K190716
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6495F
Device Catalogue Number6495F
Device Lot NumberCAB678206F
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/09/2023
Initial Date FDA Received07/06/2023
Supplement Dates Manufacturer Received07/27/2023
Supplement Dates FDA Received08/04/2023
Date Device Manufactured01/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexMale
Patient Weight89 KG
-
-