• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. LINER STANDARD 3.5 MM OFFSET 36 MM I.D. FOR USE WITH 50/52/54 MM O.D. SHELLS; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. LINER STANDARD 3.5 MM OFFSET 36 MM I.D. FOR USE WITH 50/52/54 MM O.D. SHELLS; PROSTHESIS, HIP Back to Search Results
Catalog Number 00630505036
Device Problem Device Dislodged or Dislocated (2923)
Patient Problems Failure of Implant (1924); Necrosis (1971); Joint Dislocation (2374); Ambulation Difficulties (2544)
Event Date 12/23/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2023-00061; 0002648920-2023-00149.Cat #: 00771100720 / femoral stem 12/14 neck taper plasma sprayed press-fit cementless size 7.5 extended offset / lot #: 61171199; cat #: 00801803602 / femoral head sterile product do not resterilize 12/14 taper / lot #: 61401792 the device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that a patient underwent a right hip revision approximately nine years post implantation due to chronic instability.During the procedure, it was confirmed that the patients hip joint easily dislocated.There was also extensive necrotic bone and tissue that required debulking and debridement.Fluid accumulation, corrosion, muscle damage and a large pseudotumor were found.The head and liner were removed and replaced without further complication.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: the device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
No additional event information to report at this time.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Dhr was reviewed and no discrepancies related to the reported event were found.This complaint is confirmed based on the provided medical records.Root cause is unchanged.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINER STANDARD 3.5 MM OFFSET 36 MM I.D. FOR USE WITH 50/52/54 MM O.D. SHELLS
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17271715
MDR Text Key318511897
Report Number0002648920-2023-00150
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K002960
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 08/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2014
Device Catalogue Number00630505036
Device Lot Number61354997
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/16/2022
Initial Date FDA Received07/06/2023
Supplement Dates Manufacturer Received07/12/2023
08/03/2023
Supplement Dates FDA Received07/12/2023
08/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/31/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE H10.
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexFemale
-
-