Model Number 690R32 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Tricuspid Valve Insufficiency/ Regurgitation (4453)
|
Event Date 06/06/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been returned for analysis.A supplemental report will be filed upon completion of the analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that 6 years and 6 months post implant of this 32mm tricuspid annuloplasty ring, it was explanted and replaced with a non-medtronic product. the tricuspid annuloplasty (tap) was performed concomitantly during the native mitral valve replacement (mvr), and aortic valve replacements (avr). the reason for the tricuspid annuloplasty (tap) replacement was reported as severe tricuspid regurgitation (tr), listed as a secondary cause of exacerbation by severe mitral stenosis (ms), and mitral regurgitation (mr).It was reported that the physician thought that the tricuspid ring was not the cause of the issue.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Product analysis: upon receipt at medtronic's quality laboratory, visual examination revealed that green and white multifilament sutures remain attached to the ring.A cut through the fabric was observed exposing the underlying stiffener; damage likely occurred during explant procedure.Thick, glistening off-white pannus covers the inflow aspect of the ring extending moderately to the outflow.Radiography showed traces of calcification on the ends of the ring.Radiography did not reveal a fracture.H3: device evaluated? updated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|