ABBOTT DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 7122Q/65
Device Problems Failure to Capture (1081); Retraction Problem (1536); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/30/2023
Event Type  Injury  
Event Description
During a clinic follow-up, a loss of capture was observed on the right ventricular (rv) lead.Diagnostic imaging was performed and confirmed a dislodgement of the rv lead.During the revision procedure, the helix of the rv lead was unable to be retracted.As a result, the rv lead was explanted and replaced to resolve the event.The patient was stable and will continue to be monitored.
 
Manufacturer Narrative
The reported events were dislodgement, failure to capture, and a helix mechanism issue.As received, a complete lead was returned for analysis.The reported event of helix mechanism issue was confirmed.The lead was returned with the helix retracted and clogged with blood/tissue.X-ray inspection found the inner coil over torqued at the connector region, which is consistent with procedural damage.X-ray examination of the helix mechanism found no anomalies or distortion of the helix that would have contributed to the helix mechanism issues reported in the field.The cause of the reported event was isolated to blood/tissue in the helix region and an over torqued the inner coil.The reported event of failure to capture was not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17276447
MDR Text Key318549034
Report Number2017865-2023-25222
Device Sequence Number1
Product Code NVY
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7122Q/65
Device Catalogue Number7122Q-65
Device Lot NumberA000136538
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/19/2023
Initial Date FDA Received07/06/2023
Supplement Dates Manufacturer Received07/27/2023
Supplement Dates FDA Received08/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/18/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-