ABBOTT DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES
|
Back to Search Results |
|
Model Number 7122Q/65 |
Device Problems
Failure to Capture (1081); Retraction Problem (1536); Device Dislodged or Dislocated (2923)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/30/2023 |
Event Type
Injury
|
Event Description
|
During a clinic follow-up, a loss of capture was observed on the right ventricular (rv) lead.Diagnostic imaging was performed and confirmed a dislodgement of the rv lead.During the revision procedure, the helix of the rv lead was unable to be retracted.As a result, the rv lead was explanted and replaced to resolve the event.The patient was stable and will continue to be monitored.
|
|
Manufacturer Narrative
|
The reported events were dislodgement, failure to capture, and a helix mechanism issue.As received, a complete lead was returned for analysis.The reported event of helix mechanism issue was confirmed.The lead was returned with the helix retracted and clogged with blood/tissue.X-ray inspection found the inner coil over torqued at the connector region, which is consistent with procedural damage.X-ray examination of the helix mechanism found no anomalies or distortion of the helix that would have contributed to the helix mechanism issues reported in the field.The cause of the reported event was isolated to blood/tissue in the helix region and an over torqued the inner coil.The reported event of failure to capture was not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.
|
|