RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Charging Problem (2892)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/27/2023 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint by the customer on the v60 indicating that during patient setup, the device did not show that the battery was charging and was not working on alternating current (ac) power when plugged into the outlet.It was reported that there was no patient involvement at the time the issue was discovered.The customer informed the rse that the power light emitting diode (led) was illuminated, but the battery led was not illuminated with the battery connected.The customer also stated that the device operated on battery power only, the power cord was working properly at 120vac at the power cord and ac inlet, and there was no voltage at j1 of power supply.The rse advised the customer to replace the power supply to correct the reported issue and provided the customer with the part number and price of the replacement power supply for repair.
|
|
Manufacturer Narrative
|
Multiple attempts have been made to try to obtain further information about this case, but no response was received from the customer.This file is closed and can be reopened if new information becomes available.
|
|