• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. PRUITT F3 CAROTID SHUNT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. PRUITT F3 CAROTID SHUNT Back to Search Results
Model Number 2012-10
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/07/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned for investigation.It was confirmed to be leaking at the stopcock joint.The root cause of the malfunction was due to a manufacturing operator error, not enough glue was applied on the stopcock joint during the assembly process.Due to reports of similar issues, the product design is being reviewed to address this issue.We have conducted a lot history review and did not find any issues noted in the manufacturing or packaging process that could be related to this issue.All qc tests passed their requirements.Further, we have not received any other complaints of a similar nature for devices from this lot.
 
Event Description
It was reported the pruitt f3 carotid shunt was leaking during preuse test.This device was not used for the operation.The operation was successfully completed using a replacement (hospital stock).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRUITT F3 CAROTID SHUNT
Type of Device
CAROTID SHUNT
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC
63 second ave
burlington MA 01803
Manufacturer Contact
peter song
63 second ave
burlington, MA 01803
7814251683
MDR Report Key17277436
MDR Text Key319394768
Report Number1220948-2023-00123
Device Sequence Number1
Product Code MJN
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K182916
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2012-10
Device Catalogue Number2012-10
Device Lot NumberPFT3958
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/07/2023
Initial Date FDA Received07/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-