• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number AGFN-756
Device Problems Material Frayed (1262); Shipping Damage or Problem (1570)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/18/2023
Event Type  malfunction  
Event Description
It was reported that on (b)(6) 2023, a 21mm sjm regent mechanical heart valve (flexcuff), was selected for an implant.The surgeon removed the valve from the operating room warehouse, opened the package and found that the valve ring was partially damaged.The physician did not want to waste the valve and thought that implantation would not affect the performance of the valve.The physician attempted implanting it in the patient but removed it due to difficulty suturing.A new non-abbott valve was used to complete the procedure.The product box containing the device was also noted to have been damaged.The patient is stable.
 
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
The reported event of damaged packaging and device damage was reported.The investigation confirmed that the sewing cuff was torn/frayed and contained blood.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Image from the field appeared to show the sweing cuff.Based on the information received and returned device analysis, the reported damaged packaging and device damage could not be conclusively determined.However, incident is consistent with damage during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE WITH FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key17282830
MDR Text Key318837558
Report Number2135147-2023-02939
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005975
UDI-Public05414734005975
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAGFN-756
Device Catalogue Number21AGFN-756
Device Lot Number7613715
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2023
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/15/2023
Initial Date FDA Received07/07/2023
Supplement Dates Manufacturer Received07/07/2023
Supplement Dates FDA Received07/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/05/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age52 YR
Patient SexMale
Patient Weight56 KG
-
-