Model Number 4515-70-101
Device Problems Malposition of Device (2616); Intermittent Loss of Power (4016)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 06/21/2023
Event Type  malfunction  
Event Description
It was reported that during a total knee arthroplasty surgical procedure, it was observed that while using the robotic-assisted solution satellite station device, it was observed that the saw stuttered and had cut inaccuracies on the posterior chamfer.It was reported that the cut was verified with the pointer.It was reported that the clinician did not feel comfortable using cementless components due to inaccurate cuts.It was reported that on one of the cases the sasi connection to the saw was loose.It was reported that the tensioner tends to over tension the flexion gap.It was reported that the device was being used with a robotic assisted base station device.There were no delays in the procedure.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.D10, concomitant medical devices and therapy dates, base station device, saw interface device, (b)(6) 2023.Udi: (b)(4).
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device was evaluated.The device log files were reviewed for this event and the reported condition was confirmed.The analysis of the log files could not determine a probable cause of the event since the system was performing within expected performance limits of the system.An assignable root cause could not be established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELYS SATELLITE STATION
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
DEPUY IRELAND
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthpaedic dr.
warsaw IN 46581
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17286679
MDR Text Key318782861
Report Number1818910-2023-13679
Device Sequence Number1
Product Code OLO
UDI-Device Identifier10603295519492
UDI-Public10603295519492
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K202769
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4515-70-101
Device Catalogue Number451570101
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/29/2023
Initial Date FDA Received07/10/2023
Supplement Dates Manufacturer Received08/18/2023
Supplement Dates FDA Received08/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-