• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V2
Device Problem Material Protrusion/Extrusion (2979)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/11/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation, and the customer's allegation was confirmed; due to a pinhole in the bending section cover, water tightness was lost.It was also found the metal braid protruding was due to the bending section cover being broken.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer initially reported to olympus that the uretero-reno videoscope's bending rubber had a leak.There were no reports of patient harm.The device was returned for evaluation.During the device evaluation, the following reportable malfunction was identified: the metal braid inside the bending tube was protruding.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the cause of the event is likely due to physical stress during user handling.It was possible that damaged components of the bending section stuck into bending section cover (a-rubber) during user handling, leading to damage of the a-rubber.The event can be detected by following the instructions for use (ifu) section: ifu: urf-v2/v2r operation manual chapter 3 preparation and inspection.The event can be prevented by following the ifu which state: do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.- chapter 4 operation 4.1 warnings and cautions: operation.:do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.:do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged.) olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17294949
MDR Text Key318932830
Report Number9610595-2023-09918
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170343582
UDI-Public04953170343582
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K172246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/14/2023
Initial Date FDA Received07/10/2023
Supplement Dates Manufacturer Received07/26/2023
Supplement Dates FDA Received08/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/17/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-