Catalog Number 175816 |
Device Problem
Dent in Material (2526)
|
Patient Problem
Discomfort (2330)
|
Event Date 06/20/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
|
Event Description
|
It was reported that there was an issue on at least 2 occasions where the foley catheter balloon did not inflate properly.One fell out of a patient, the other was a test (not used) where the balloon inflated however the balloon had a hole and leaked.Product was deformed, foley catheter balloon filled up with 10cc was lopsided.Per follow up via email on 28jun2023, it was stated that the only sample available was the foley that had a hole in the tubing, not the foley where the balloon was lopsided because it was used, they could not save it.Patient with lopsided foley was okay, but was very uncomfortable therefore they removed the foley and that was when they realized that the product was deformed.
|
|
Event Description
|
It was reported that there was an issue on at least 2 occasions where the foley catheter balloon did not inflate properly.One fell out of a patient, the other was a test (not used) where the balloon inflated however the balloon had a hole and leaked.Product was deformed, foley catheter balloon filled up with 10cc was lopsided.Per follow up via email on 28jun2023, it was stated that the only sample available was the foley that had a hole in the tubing, not the foley where the balloon was lopsided because it was used, they could not save it.Patient with lopsided foley was okay, but was very uncomfortable therefore they removed the foley and that was when they realized that the product was deformed.
|
|
Manufacturer Narrative
|
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be "balloon molding".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "sterile unless package is opened or damaged, except for any individually packaged components within the tray which are not labeled as sterile.These components are not terminally sterilized.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Single use only.Do not resterilize.For urological use only.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter and may cause balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Please consult product label and insert for any indications, contraindications, hazards, warnings, cautions and directions for use.Proper techniques for urinary catheter insertion ¿ perform hand hygiene immediately before and after insertion ¿ insert urinary catheters using aseptic technique and sterile equipment ¿ use the smallest foley catheter possible, consistent with good drainage ¿ document the indications for catheter insertion, date and time of catheter insertion, individual who inserted catheter, and date and time of catheter removal in patient record proper techniques for urinary catheter maintenance ¿ secure the foley catheter, use the statlock® foley stabilization device if provided ¿ maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions ¿ maintain unobstructed urine flow and keep the catheter and collection tube free from kinking ¿ keep the collection bag below the level of the bladder or hips at all times ¿ empty the collection bag regularly (e.G., prior to transport) using a separate, clean collection container for each patient ¿ routine hygiene (e.G., cleansing of the meatal surface during daily bathing or showering) is appropriate ¿ leave foley catheter in place only as long as needed.Bard® ez-lok® sampling port (indicated by the blue stem in the port) accepts luer lock (fig.1a) or slip tip syringes (fig.1b).1.Occlude drainage tubing a minimum of 3 inches below the sampling port by kinking the tubing until urine is visible under the access site.2.Swab surface of bard® ez-lok® sampling port with antiseptic wipe.(fig.2) 3.Using aseptic technique, position the syringe in the center of the sampling port.Press the syringe firmly and twist gently to access the sampling port.(fig.3) 4.Slowly aspirate urine sample into syringe and remove syringe from sample port.5.Unkink tubing if necessary and transfer urine specimen into specimen cup or follow hospital protocol.Discard syringe according to hospital protocol.6.Follow established hospital protocol for specimen labeling and transport to lab." the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|