MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPT; Stimulator, electrical, implanted, for parkinsonian tremor Back to Search Results
Model Number B35200
Device Problem Electromagnetic Interference (1194)
Patient Problems Shaking/Tremors (2515); Insufficient Information (4580)
Event Date 06/02/2023
Event Type  malfunction  
Event Description
It was reported that the patient had an mri of their lumbar spine in the last couple of weeks and the results came back "cloudy" because the patient could not control their movement/tremors due to the implantable neurostimulator (ins) being in mri mode.The patient noted that their tremors were more profound and the mri exacerbated the tremor in their lower lumbar.As a result, the patient's surgeon could not verify what they needed to look at, and they may need to do an mri while the patient is under anesthesia.Agent reviewed anesthesia compatibility information.The patient confirmed that their movement/tremors were under control once they exited out of mri mode.Agent advised the patient to have their healthcare provider (hcp) call if they have questions regarding mri preparation.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date.The report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEPT
Type of Device
Stimulator, electrical, implanted, for parkinsonian tremor
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key17297776
MDR Text Key319696537
Report Number3004209178-2023-12421
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00763000420987
UDI-Public00763000420987
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/14/2023
Device Model NumberB35200
Device Catalogue NumberB35200
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/16/2023
Initial Date FDA Received07/11/2023
Date Device Manufactured04/24/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
-
-