Model Number 97715
Device Problems Failure to Deliver Energy (1211); Energy Output Problem (1431); Difficult to Remove (1528); Application Program Problem (2880)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/24/2023
Event Type  Injury  
Event Description
Information was received from a manufacturer representative regarding a patient who was implanted with an implantable neurostimulator (ins) for non-malignant pain.Caller is with pt today and notes the pt's stimulation got weaker and 'just stopped' without any event/issue prompting this.The pt had checked to confirm it was on, pt charged the ins.The caller checked impedances today and they were all 'green' with a broad range of 620-3000ohms.The caller noted he tried programming on all 4 available electrodes at 300usec 80hz with bipoles, double guarded cathodes, etc, but the pt only felt stimulation 'for a moment but it went away'.The caller notes getting oor around 6ma when turning stim up.The caller will look to do imaging at this time.(b)(6)2023: additional information was received from a manufacturer representative (rep).The rep reported that the cause of the out of regulation message and stimulation just stopping was not determined.Rep tried different electrode configurations and adjusted settings.Impedances were all normal and in range.The issue is not resolved.The provided information has been confirmed with the physician/account.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received on september 8, 2023.The rep reported that the pt was going to have the system replaced on (b)(6) 2023.During the explant procedure the surgeon had difficulty accessing and removing the paddle lead, so they decided to only do the explant portion on (b)(6) 2023.The replacement surgery will occur soon (date of replacement surgery was not provided).
 
Manufacturer Narrative
Continuation of d10: product id 3587a lot# l58159 implanted: (b)(6) 1999 explanted: (b)(6) 2023 product type lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Continuation of d10: product id 3587a lot# l58159 serial# implanted: (b)(6) 1999, explanted: (b)(6) 2023, product type lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Device was discarded.Oor message resolved by lowering intensity.Unable to obtain patient weight.Physician was notified and was there of event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key17299375
MDR Text Key319421391
Report Number3004209178-2023-12447
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169781702
UDI-Public00643169781702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 10/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/28/2020
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/27/2023
Initial Date FDA Received07/11/2023
Supplement Dates Manufacturer Received09/08/2023
10/05/2023
Supplement Dates FDA Received10/05/2023
10/24/2023
Date Device Manufactured08/05/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
"SEE H10...."; "SEE H10...."
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient SexMale
-
-