G4 ¿ pma/510(k) #: exempt.This report includes information known at this time.A follow-up report will be submitted should additional, relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
Investigation / evaluation: on 06jul2023, cook medical inc.Received a complaint from the vascular international physicians facility, located in the city of indianapolis in.It was reported that upon the placement of a ultrathane mac-loc locking loop multipurpose drainage catheter (rpn: ult10.2-38-45-p-6s-clm-rh; lot: 15423166 ), the stiffening cannula was difficult to remove from the drainage catheter.Another like device was used to complete the procedure successfully.No adverse effects have been reported due to this occurrence.Reviews of documentation including the complaint history, device history record (dhr), quality control procedures, and instructions for use (ifu) of the complaint device were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhrs for the reported complaint device lot: 15423166 and the related subassembly lots revealed no relevant non-conformances.To date, a further search of our database records revealed no other complaints with the reported lot number.Cook also reviewed product labeling.The product ifu, [t_multi2] ¿multipurpose drainage catheter,¿ provides the following information to the user related to the reported failure mode: precautions: ¿when inserting a stiffening cannula into a catheter with retention suture, hold suture during cannula insertion to avoid bunching or tangling of suture.¿ how supplied: ¿supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened or undamaged.Do not upset he product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.¿ the information provided upon review of the dmr and ifu suggests that the device was not manufactured out of specification, and that there are no nonconforming devices in house or out in the field.Based on the information provided, no product returned, and the results of the investigation, cook has concluded that component failure without a manufacturing or design defect contributed to this incident.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|