Model Number 7122Q/58
Device Problem Retraction Problem (1536)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/26/2023
Event Type  malfunction  
Event Description
It was reported during implant procedure that the helix of the right ventricular lead exhibited a failure to retract.The lead was exchanged.There were no patient consequences.
 
Manufacturer Narrative
The reported event of helix mechanism issue was confirmed.As received, a complete lead was returned in one piece.The lead was returned with the helix stretched, bent, and clogged with dried blood.X-ray examination found the helix to be stretched and bent at the helix region consistent with procedural damage.The cause of helix mechanism issue was isolated to the helix being stretched, bent, and clogged with dried blood.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17303082
MDR Text Key319413818
Report Number2017865-2023-35630
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734503174
UDI-Public05414734503174
Combination Product (y/n)Y
Reporter Country CodeTW
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7122Q/58
Device Catalogue Number7122Q-58
Device Lot NumberA000139875
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/26/2023
Initial Date FDA Received07/11/2023
Supplement Dates Manufacturer Received07/24/2023
Supplement Dates FDA Received07/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-