MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SPYSCOPE DS II ACCESS & DELIVERY CATHETER; CHOLEDOCHOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SPYSCOPE DS II ACCESS & DELIVERY CATHETER; CHOLEDOCHOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID Back to Search Results
Model Number M00546610
Device Problems Poor Quality Image (1408); Optical Problem (3001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2023
Event Type  malfunction  
Manufacturer Narrative
Block e1: initial reporter phone: (b)(6).Block h6: imdrf impact code f1001 is being used to capture the reportable event of aborted/cancelled procedure.
 
Event Description
It was reported to boston scientific corporation that a spyscope ds ii access & delivery catheter was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6), 2023.During the procedure, there was a sort of interference and partial loss of visualization of the image on the screen.Some minutes later, the screen did not show any image.They unplugged and replugged the spyscope a couple of times but the problem persisted.The procedure was not completed due to this event.There were no patient complications reported as a result of this event.
 
Manufacturer Narrative
Block e1: initial reporter phone: (b)(6).Block h6: imdrf impact code f1001 is being used to capture the reportable event of aborted/cancelled procedure.Block h10 the returned spyscope ds ii was analyzed, and a visual evaluation noted that there were elevator marks on the shaft of the catheter.An image assessment for visualization was performed.Upon plugging the device into the controller, a live image was displayed.Articulation of the catheter had no effect on the image.X-ray imaging of the distal tip showed no problem with the redistribution layer (rdl).No camera wire damage was observed near the distal end.No camera wire damage was observed in the pebax region of the catheter.X-ray imaging of the handle showed no problem with the breakout region or camera wires around the strain relief.X-ray imaging shows no damage to the camera wires at the printed circuit board assembly (pcba).The handle was opened and the electrical components inside were inspected visually.There was procedural residue seen in the plastic optic fibers (pof), visual assessment showed no problem with the camera wires in the glue feature.The bond of the glue feature to the pcba was inspected, tweezers were used to wiggle the glue feature.The connection of the camera wires to the pcba was also inspected by slightly lifting the bottom of the glue feature for each of the four camera wires using the tip of the tweezers.No impact to image was seen after these interactions.A saline test was conducted, based on residue at the pofs, and after insertion of saline, enough to reach the handle, the image was disrupted.Purple and pink lines appeared on the screen followed by the initialization screen.The reported complaint regarding visualization was confirmed.During product analysis, a live image was seen after insertion of device into controller.After a saline test, image was disrupted.Purple and pink lines appeared followed by the initialization screen.During procedure, fluid likely back flowed into the optics lumen and caused a change in capacitance of the electrical circuit resulting in a loss of image.An investigation to address this problem has been completed.Based on all gathered information, the probable cause selected for the image failure is cause traced to device design, which indicates that the problems are traced to the design specifications.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event Description
It was reported to boston scientific corporation that a spyscope ds ii access & delivery catheter was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2023.During the procedure, there was a sort of interference and partial loss of visualization of the image on the screen.Some minutes later, the screen did not show any image.They unplugged and replugged the spyscope a couple of times but the problem persisted.The procedure was not completed due to this event.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPYSCOPE DS II ACCESS & DELIVERY CATHETER
Type of Device
CHOLEDOCHOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17309598
MDR Text Key319670323
Report Number3005099803-2023-03565
Device Sequence Number1
Product Code FBN
UDI-Device Identifier08714729965404
UDI-Public08714729965404
Combination Product (y/n)N
Reporter Country CodeCS
PMA/PMN Number
K183636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/15/2024
Device Model NumberM00546610
Device Catalogue Number4661
Device Lot Number0029052767
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/15/2023
Initial Date FDA Received07/12/2023
Supplement Dates Manufacturer Received08/01/2023
Supplement Dates FDA Received08/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-