Model Number N/A |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Scar Tissue (2060)
|
Event Date 09/16/2022 |
Event Type
Injury
|
Event Description
|
A journal article was retrieved from journal of surgical oncology (2022) that reported a study from the west coast of the united states.The study reported 20 patients experienced type 1a soft tissue functional failures which consisted exclusively of arthrofibrosis.Patients were treated with manipulation under anesthesia, lysis of adhesions, chemodenervation, and/or revision of the implant.Attempts have been made and no further information has been provided.
|
|
Manufacturer Narrative
|
(b)(4).G2 - literature: tanaka ks, andaya vr, thorpe sw, et al.Survival and failure modes of the compress® spindle and expandable distal femur endoprosthesis among pediatric patients: a multi-institutional study. j surg oncol.2023;127(1):148-158.Doi:10.1002/jso.27094.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Arthrofibrosis is defined as the development of fibrous scar tissue within or surrounding a joint.Arthrofibrosis is a known postoperative procedure related complication that can occur from surgical implantation of new joint replacement as well as from previous injuries or surgical procedures.Scar tissue formation is a normal healing response; however, the buildup of such can result in pain, stiffness, limited range of motion, and difficulty properly ambulating.If excess scar tissue develops, conservative measures such as exercises or physical therapy would be attempted first.If these attempts fail, surgical intervention such as manipulation under anesthesia, arthroscopic arthrolysis, or open arthrotomy would become necessary to remove the fibrous tissue and restore joint function.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Search Alerts/Recalls
|
|
|