• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 5.0 MM BARREL BURR PLUS 5PK; ARTHROSCOPIC SHAVER SYSTEM BLADE, SINGLE-USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 5.0 MM BARREL BURR PLUS 5PK; ARTHROSCOPIC SHAVER SYSTEM BLADE, SINGLE-USE Back to Search Results
Model Number 283869
Device Problems Device Contaminated During Manufacture or Shipping (2969); Mechanical Jam (2983); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: udi: (b)(4).E3: reporter is a j&j sales representative.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported by the sales rep in switzerland that during an unknown procedure on an unknown date, it was observed that the 5.0 mm barrel burr plus device was blocked.It was further reported that when separating inner part from outer part.They had pieces that rubbed off.Another like device was used to complete the procedure with a delay of five minutes.There were no adverse patient consequences reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary
=
> the device was received and evaluated.Upon visual inspection, the inner and outer sleeves were separated and inspected for any visual defects that may contribute to the complaint condition.The device shows friction marks on the surface of the distal end of the outer blade.Also, the inner blade showed points of wear near the distal part.A manufacturing record evaluation was performed for the finished device m2206003 number, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection result, this complaint can be confirmed.The possible root cause for reported failure can be attributed to procedural variables, such handling of the device or product interaction during procedure; as per ifu; excessive side loading may result in blade wear and degradation.Adequate suction is recommended to reduce wear and degradation of the device.Besides, a hand piece used in this procedure where the blade is assembled was not received for evaluation, potential ways to reduce shedding is to ensure all hand pieces are properly serviced and maintained.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0 MM BARREL BURR PLUS 5PK
Type of Device
ARTHROSCOPIC SHAVER SYSTEM BLADE, SINGLE-USE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key17321574
MDR Text Key319252374
Report Number1221934-2023-02748
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705023080
UDI-Public10886705023080
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K041824
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283869
Device Catalogue Number283869
Device Lot NumberM2206003
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/03/2023
Initial Date FDA Received07/13/2023
Supplement Dates Manufacturer Received08/10/2023
08/17/2023
Supplement Dates FDA Received08/11/2023
08/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-