Model Number 862134
Device Problem Erratic Results (4059)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/23/2023
Event Type  malfunction  
Event Description
The customer reported that the patient was sent to the emergency department for rescue due to "cardiopulmonary arrest for 20 minutes" from 120, and electrocardiogram monitoring was given according to the doctor's advice.After cardiopulmonary resuscitation for more than 10 minutes, the patient recovered his autonomous heart rate, and the ecg waveform of the electrocardiogram monitor did not show.The ecg waveform still showed a straight line, which was not consistent with the reality.Replace the ecg monitor immediately and ask the maintenance personnel of the equipment department for maintenance.
 
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.
 
Manufacturer Narrative
A philips post market surveillance manager (pmsm) remotely interviewed the customer and confirmed there was no issues with the 5 lead ecg electrodes, and they were purchased from philips.The root cause of the issue was determined to be the parameter board.Once the parameter board was replaced, the unit was returned to user specification.After parameter board replacement the device was returned to functional use with no further issues identified.The device remains at the customer site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MP20
Type of Device
INTELLIVUE MP20
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key17321702
MDR Text Key319346283
Report Number9610816-2023-00351
Device Sequence Number1
Product Code MHX
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K122439
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number862134
Device Catalogue Number862134
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/19/2023
Initial Date FDA Received07/13/2023
Supplement Dates Manufacturer Received06/28/2023
Supplement Dates FDA Received07/24/2023
Date Device Manufactured09/19/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-