• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ELEKTA INC MOSAIQ; ACCELERATOR, LINEAR, MEDICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ELEKTA INC MOSAIQ; ACCELERATOR, LINEAR, MEDICAL Back to Search Results
Model Number MOSAIQ
Device Problem Computer Software Problem (1112)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/11/2023
Event Type  malfunction  
Event Description
The customer reported an instance of incorrect dose recording.
 
Manufacturer Narrative
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
 
Manufacturer Narrative
The investigation was completed by conducting a thorough evaluation of the product and the reported information.The customer reported an instance of incorrect dose recording.From the mosaiq logs the investigation found that the sequencer crashed mid treatment, so the treatment was not recorded.Mosaiq received the sro file, dicom rt image for port during integrate option as well as the dicom beam records for two fields which were treated in full.However, these files were not processed because of the vmi's error, "vmicheckerrormsg: failed to invoke c-store-rq.Reason: opening association failed with impac_dcm_scp".The error indicated that there were some problems communicating with the dcmwin app.Based on available information there was no mistreatment.Mosaiq did not have any malfunction and was working as designed and intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIQ
Type of Device
ACCELERATOR, LINEAR, MEDICAL
Manufacturer (Section D)
ELEKTA INC
100 mathilda place
5th floor
sunnyvale CA 94086
Manufacturer (Section G)
ELEKTA INC
400 perimeter center terrace
suite 50
atlanta GA 30346
Manufacturer Contact
cornerstone
london road
crawley, west sussex RH10 -9BL
MDR Report Key17323228
MDR Text Key319755612
Report Number2950347-2023-00006
Device Sequence Number1
Product Code IYE
UDI-Device Identifier00858164002091
UDI-Public(01)00858164002091(10)2.64.235
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K141572
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberMOSAIQ
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/16/2023
Initial Date FDA Received07/14/2023
Supplement Dates Manufacturer Received06/16/2023
Supplement Dates FDA Received11/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/21/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-