The investigation was completed by conducting a thorough evaluation of the product and the reported information.The customer reported an instance of incorrect dose recording.From the mosaiq logs the investigation found that the sequencer crashed mid treatment, so the treatment was not recorded.Mosaiq received the sro file, dicom rt image for port during integrate option as well as the dicom beam records for two fields which were treated in full.However, these files were not processed because of the vmi's error, "vmicheckerrormsg: failed to invoke c-store-rq.Reason: opening association failed with impac_dcm_scp".The error indicated that there were some problems communicating with the dcmwin app.Based on available information there was no mistreatment.Mosaiq did not have any malfunction and was working as designed and intended.
|