COMPANION MEDICAL INC INPEN MMT-105ELBLNA ELI LILY BLUE; SYRINGE, PISTON Back to Search Results
Model Number MMT-105ELBLNA
Device Problems Break (1069); Material Too Rigid or Stiff (1544); Connection Problem (2900); Patient Device Interaction Problem (4001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/27/2023
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Information received by medtronic indicated that the customer reported the inpen dose button was difficult to press.Troubleshooting was performed and found that the inpen was damaged and the cartridge holder was also damaged.No harm requiring medical intervention was reported.The customer will discontinue using the inpen.The inpen will be returned for analysis.
 
Manufacturer Narrative
Inpen did not pair with commercial app.Inpen did not transmit to manufacturing app.Performed encoder bond investigation and found that the encoder pattern wheel tabs rotating and traveling off the keyed slots of dose nut guides.Encoder pattern wheel should never rotate.This causes an unexpected travel of the encoder pattern wheel creating resistance to dial and dispensing.Also, plastic shavings contamination builds up found caused by encoder wheel tabs rubbing at the walls of the dose nut.Abrasions down the length of the dose nut.Unable to perform baseline/wireless functionality and displacement dose accuracy test.Inpen passed front cap investigation.In conclusion: it was determined from destructive analysis that the difficult to dial/dose was caused by a pattern wheel misalignment due to an encoder base bond failure.This can affect insulin delivery.Therefore, the customer concern of dose dial being difficult to dial/dose was confirmed.Cosmetic damage was not confirmed.Cartridge holder damage was not confirmed.Unable to confirm dialing mis-alignment due to difficult to dial/dose.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105ELBLNA ELI LILY BLUE
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
justin ellis
1223 world trade drive, 100
san diego, CA 92128
7635265677
MDR Report Key17323307
MDR Text Key319108730
Report Number3012822846-2023-01287
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201337
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-105ELBLNA
Device Catalogue NumberMMT-105ELBLNA
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 06/27/2023
Initial Date FDA Received07/14/2023
Supplement Dates Manufacturer Received07/28/2023
Supplement Dates FDA Received08/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Age14 YR
Patient SexFemale
Patient Weight100 KG
-
-