• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG SOLUTIONS S.A. MAHURKAR ELITE; CATHETER, HEMODIALYSIS, NON-IMPLANTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG SOLUTIONS S.A. MAHURKAR ELITE; CATHETER, HEMODIALYSIS, NON-IMPLANTED Back to Search Results
Model Number 8888211424
Device Problems Defective Component (2292); Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/23/2023
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, on the back table before any insertion into patient, the catheter was tested for initial flush which did not go through the tip of both red and blue lumens.The catheter would not flush.There was no lumen exit site on distal tip of catheter visualized.On june 28, 2023, the rep looked at the catheter through the bag where it was placed in and noticed there were no distal slots on catheter.The catheter was not properly resembling any distal side slots which internally could be a cause for no proper flushing.No visible cracks, cuts, or holes noted.The catheter was not repaired.There was no leak.Tego was not utilized.There was no luer adapter issue.There was no dimension issue noted.The dimension of the catheter was what was indicated on the label.The customer did not try to reverse the lines.A new vascular catheter pack of the same reference number and lot number was obtained on the same day for the patient without any issue and the procedure was completed.There was no thrombosis.There was no reported patient injury.
 
Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted that the distal side slots were missing from the cannula.Flushing through the arterial luer adapter was therefore impossible.It was reported that the device was occluded and there was an issue with the catheter dimension.The reported issues were confirmed.The most likely cause was determined to be manufacturing related.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAHURKAR ELITE
Type of Device
CATHETER, HEMODIALYSIS, NON-IMPLANTED
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
CS  20101
Manufacturer (Section G)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
CS   20101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key17325198
MDR Text Key319252630
Report Number3009211636-2023-00205
Device Sequence Number1
Product Code MPB
UDI-Device Identifier10884521788459
UDI-Public10884521788459
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120674
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/20/2024
Device Model Number8888211424
Device Catalogue Number8888211424
Device Lot Number2226500218
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/28/2023
Initial Date FDA Received07/14/2023
Supplement Dates Manufacturer Received01/12/2024
Supplement Dates FDA Received01/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-