Establishment name: (b)(6) hospital.The device was returned to olympus for inspection, and the customer's issue of ¿had an image defect (color becomes purple)¿ was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 6 years since the subject device was manufactured.Based on the results of the investigation, it is likely that the following led to the malfunction: we presume that the defect was caused due to breakage of image sensor unit including disconnection by stress of repeated use, external factors, or handling, or that the components including ic chip and capacitor, mounted on the electric circuit board had a defect.The event can be prevented by following the instructions for use which state: inspection of the endoscopic system¿ describes the following warning.Confirm that the white light imaging (wli) and narrow band imaging (nbi) endoscopic images are normal.1.Before inspection, wipe the objective lens using clean lint-free cloths moistened with saline solution or sterilized water.2.Observe the palm of your hand in the wli and nbi endoscopic images.3.Confirm that light is output from the endoscope¿s distal end.4.Adjust the brightness level as appropriate.5.Confirm that the wli and nbi endoscopic images are free from noise, blur, fog, or other irregularities.6.Turn the angulation control levers slowly in each direction until it stops.7.Confirm that the wli and nbi endoscopic images do not momentarily disappear or display any other irregularities.In addition, the following non-reportable malfunctions were found during the device evaluation: due to damage on charged coupled device (ccd) unit, the image has roughness, several scratches found throughout the device, adhesive on bending section has a chip, due to damage on lock engagement lever, bending section cannot be fixed firmly, due to damage on ccd unit, the image has white dots, and video connector case has a crack.Olympus will continue to monitor field performance for this device.
|