• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. FORCEPS/IRRIGATION PLUG (ISOLATED TYPE); FORCEPS/ IRRIGATION PLUG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. FORCEPS/IRRIGATION PLUG (ISOLATED TYPE); FORCEPS/ IRRIGATION PLUG Back to Search Results
Model Number MAJ-891
Device Problems Melted (1385); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/18/2023
Event Type  malfunction  
Manufacturer Narrative
The serial number of the device has not been provided at this time.The subject device has been returned to olympus for evaluation and the reported issue was confirmed.A review of the device history record (dhr) could not be performed due to the serial number being unknown.Based on the results of the investigation, it is likely that, the reported issue was due to problem on chemicals in manual cleaning, such as the use of chemicals at high concentrations without the correct dilution conditions of the chemicals and failure of autoclave with chemical residue.The root cause could not be identified.Olympus will continue to monitor the field performance of this device.
 
Event Description
The customer reported to olympus, during reprocessing, when the forceps/irrigation plug (isolated type) was sterilized in an autoclave (132 degrees, 5 minutes) under the sterilization conditions specified in the instruction manual, the shaft of the cock melted in such a way that it stuck to the sterilization pack.There was no harm or user injury reported due to the event.Patient identifier: (b)(6) and (b)(6) capture related events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORCEPS/IRRIGATION PLUG (ISOLATED TYPE)
Type of Device
FORCEPS/ IRRIGATION PLUG
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17332196
MDR Text Key320023860
Report Number9610595-2023-10174
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170063114
UDI-Public04953170063114
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K912120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMAJ-891
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/19/2023
Initial Date FDA Received07/17/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-