ZOLL MEDICAL CORPORATION UNI-VENT, MODEL 731; VENTILATOR Back to Search Results
Model Number 799-EMBP-05-01
Device Problems Communication or Transmission Problem (2896); Failure of Device to Self-Test (2937)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/23/2023
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that while attempting to treat a (b)(6) male patient, the device displayed "5v self check failure - 1172", "internal comm failure - 1173" and "off set self check failure- 1174" error messages.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation.The customer's report of "5v self- check failure-1172" was observed during review of device data logs.However, the device was put through extensive testing including stress and troubleshoot testing without duplicating the report.The reports of "internal comm failure - 1173" and "off set self check failure - 1174" were unsubstantiated.No evidence was found within the logs.The report of no video display was not replicated or confirmed.A system interconnect ribbon cable was replaced as a precaution.The device was recertified and returned to customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNI-VENT, MODEL 731
Type of Device
VENTILATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key17334866
MDR Text Key319382889
Report Number1220908-2023-02752
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00847946024185
UDI-Public00847946024185
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number799-EMBP-05-01
Device Catalogue Number799-EMBP-05-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/26/2023
Initial Date FDA Received07/17/2023
Supplement Dates Manufacturer Received06/26/2023
Supplement Dates FDA Received08/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age27 YR
Patient SexMale
-
-