The initial reporter stated they received discrepant results for two patient samples tested with elecsys troponin t hs stat on a cobas e411 disk.No questionable results were reported outside of the laboratory.The first patient sample initially resulted in a troponin t value of 4706 ng/l.The sample was repeated twice, resulting in values of 26 ng/l and 4856 ng/l.The values of 4706 ng/l and 4856 ng/l were deemed correct.The second patient sample initially resulted in a troponin t value of 15 ng/l on (b)(6) 2023.The sample was repeated on (b)(6) 2023, resulting in a value of 41 ng/l.The repeat value was deemed correct.
|
The serial number of the cobas e411 disk is (b)(6).The field service engineer checked the system.Probe tubing, the measuring cell, syringe seals, and tubing seals were replaced.The probes and mixer were cleaned and adjusted.The photomultiplier tube high voltage was adjusted.The system volume was checked, a blank cell calibration was performed, and performance testing was run.Calibration and controls were run.
|