Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A nurse reported that during a cataract surgery while using a an ophthalmic console there was a burst of fluid and the patient experienced capsular tear also the lens fragment fell posteriorly and retained lens iridocyclitis.The system has also exhibited intra ocular pressure issue (iop).The surgery was completed and the patient is being monitored under medication.
|
|
Manufacturer Narrative
|
The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
Additional information received clarified that the patient is doing well.
|
|
Manufacturer Narrative
|
The company representative was unable to confirm nor replicate the reported event.The system was tested and found to meet product specifications.A non-conformance-based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.The system was found to meet specifications.Therefore, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|