Device evaluated by mfr.: visual inspection was performed.The main coil, the pusher wire and the introducer sheath returned for the analysis.The main coil was stuck inside the introducer sheath, and it was necessary to make cuts on the introducer to free the coil.On the device it was observed that the main coil stretched and kinked.Additionally, it was detached at the coil arm section.No more damages were observed.Under the microscope it was observed the main coil stretched and kinked.Additionally, it was detached at the coil arm section.The functional inspection could not be performed because the main coil and the pusher wire are not interlocking.Dimensional inspection was performed, and the zap tip and primary coil outer diameter passed the test.
|
Reportable based on device analysis completed on (b)(6) 2023.It was reported that advancing difficulties were encountered.The target lesion was located in the splenic artery.A 8mm x 20cm f-idc 2d coil was selected for use.During the procedure, the coil was advanced into the middle part of the delivery system but could no longer be advanced.The coil was removed and the procedure was completed with a different device.No patient complications were reported.However, returned device analysis revealed a detached coil at the arm section.
|