Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: b5 this report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation ¿ evaluation: it was reported that leakage was observed at the mac-loc adaptor of the ultrathane dawson-mueller mac-loc locking loop multipurpose drainage catheter.A new drainage catheter was placed with no adverse effects to the patient.The patient did not have tortuous or calcified anatomy.Reviews of documentation including the complaint history, quality control procedures, manufacturing instructions (mi), and instructions for use (ifu) of the complaint device were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded sufficient inspection activities are in place to identify this failure mode prior to distribution.The customer did not provide the lot number for the complaint device.Cook reviewed the sales history for this customer and was unable to identify the complaint lot.The device history record could not be reviewed.Based on this information, the device was manufactured to specification.There is no evidence of nonconforming material in house or in the field.Cook reviewed the product labeling for the complaint device.The instructions for use (ifu) [ t_multi2_rev1, ultrathane dawson-mueller mac-loc locking loop multipurpose drainage catheter] states the following: how supplied: "upon removal from package, inspect the product to ensure no damage has occurred." instructions for use: for locking loop catheters, lock the catheter in place using appropriate technique for the locking mechanism type, as described below.For mac-loc locking loop mechanism: a.Stabilize the mack-loc catheter hub assembly with one hand and pull back on the drawstring to form the distal catheter loop configuration.B.While maintaining traction on the drawstring, push the locking cam lever down until a distinct ¿snap¿ is felt.The distal loop of the catheter is now locked into position.C.Trim off the excess drawstring.Based on the information provided, no product returned, and the results of the investigation, cook medical has concluded that component failure unrelated to manufacturing or design deficiencies contributed to this incident.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|