Model Number 1028772 |
Device Problem
Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/23/2023 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint by the customer on the bi-pap indicating that the device shut off when it is unplugged.The device was in use on a patient at the time the reported issue was discovered; however, there was no harm reported to the patient or user.The manufacturer's field service engineer (fse) was dispatched to the site and confirmed the reported problem.Device evaluation and repair are pending.Investigation is ongoing.
|
|
Manufacturer Narrative
|
The manufacturer's field service engineer (fse) was dispatched to the site and confirmed the reported problem.The fse stated that they were told that the device was not stored properly.The fse stated that they sought the opinion of others and that, together, they determined that the device was more than likely not stored properly.Furthermore, it is probable that the device was stored plugged in, with the power switch turned on which depleted the internal battery over time.Multiple attempts have been made to try to obtain further information about this case but no response received from the customer.This file is closed and can be reopened if new information becomes available.
|
|