RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Model Number 1028772
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/23/2023
Event Type  malfunction  
Event Description
Philips received a complaint by the customer on the bi-pap indicating that the device shut off when it is unplugged.The device was in use on a patient at the time the reported issue was discovered; however, there was no harm reported to the patient or user.The manufacturer's field service engineer (fse) was dispatched to the site and confirmed the reported problem.Device evaluation and repair are pending.Investigation is ongoing.
 
Manufacturer Narrative
The manufacturer's field service engineer (fse) was dispatched to the site and confirmed the reported problem.The fse stated that they were told that the device was not stored properly.The fse stated that they sought the opinion of others and that, together, they determined that the device was more than likely not stored properly.Furthermore, it is probable that the device was stored plugged in, with the power switch turned on which depleted the internal battery over time.Multiple attempts have been made to try to obtain further information about this case but no response received from the customer.This file is closed and can be reopened if new information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key17351638
MDR Text Key319430827
Report Number2518422-2023-16183
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K053168
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1028772
Device Catalogue Number1028772
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/23/2023
Initial Date FDA Received07/19/2023
Supplement Dates Manufacturer Received09/06/2023
Supplement Dates FDA Received09/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/06/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-