Catalog Number 8607000
Device Problems Fire (1245); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/19/2023
Event Type  Injury  
Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.
 
Event Description
Device motor burned out during surgery, affecting the progress of the surgery.
 
Event Description
Device motor burned out during surgery, affecting the progress of the surgery.
 
Manufacturer Narrative
The investigation was performed based on the available information.The service was performed by thirty party and dräger was not contacted.No additional information, no parts and no log was available.Thus, no root cause could be determined.The motor itself is enclosed in metal parts that are non-flammable.No comparable complaints are known.It is recommended to have the fabius serviced regularly by trained personnel.H3 other text : device not available for investigation, 3rd party service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FABIUS GS PREMIUM
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
Manufacturer Contact
moislinger allee 53-55
lübeck 23542
4518822868
MDR Report Key17359209
MDR Text Key319370010
Report Number9611500-2023-00268
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K011404
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8607000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/17/2023
Initial Date FDA Received07/20/2023
Supplement Dates Manufacturer Received08/07/2023
Supplement Dates FDA Received09/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-